Dosimetric effect of limited aperture multileaf collimator on VMAT plan quality: A study of prostate and head-and-neck cancers

dc.contributor.authorMurtaza, Ghulam
dc.contributor.authorMehmood, Shahid
dc.contributor.authorRasul, Shahid
dc.contributor.authorMurtaza, Imran
dc.contributor.authorEhsan ullah Khan
dc.date.accessioned2019-03-18T15:10:49Z
dc.date.available2019-03-18T15:10:49Z
dc.date.issued2018-02-16
dc.description.abstractThe aim of study was to evaluate the dosimetric effect of collimator-rotation on VMAT plan quality, when using limited aperture multileaf collimator of Elekta Beam Modulator™ providing a maximum aperture of 21 cm × 16 cm. Background The increased use of VMAT technique to deliver IMRT from conventional to very specialized treatments present a challenge in plan optimization. In this study VMAT plans were optimized for prostate and head and neck cancers using Elekta Beam-ModulatorTM, whereas previous studies were reported for conventional Linac aperture. Materials and methods VMAT plans for nine of each prostate and head-and-neck cancer patients were produced using the 6 MV photon beam for Elekta-SynergyS® Linac using Pinnacle3 treatment planning system. Single arc, dual arc and two combined independent-single arcs were optimized for collimator angles (C) 0°, 90° and 0°–90° (0°–90°; i.e. the first-arc was assigned C0° and second-arc was assigned C90°). A treatment plan comparison was performed among C0°, C90° and C(0°–90°) for single-arc dual-arc and two independent-single-arcs VMAT techniques to evaluate the influence of extreme collimator rotations (C0° and 90°) on VMAT plan quality. Plan evaluation criteria included the target coverage, conformity index, homogeneity index and doses to organs at risk. A ‘two-sided student t-test’ (p ≤ 0.05) was used to determine if there was a significant difference in dose volume indices of plans. Results For both prostate and head-and-neck, plan quality at collimator angles C0° and C(0°–90°) was clinically acceptable for all VMAT-techniques, except SA for head-and-neck. Poorer target coverage, higher normal tissue doses and significant p-values were observed for collimator angle 90° when compared with C0° and C(0°–90°). Conclusions A collimator rotation of 0° provided significantly better target coverage and sparing of organs-at-risk than a collimator rotation of 90° for all VMAT techniques.en_US
dc.identifier.citationMurtaza, G., Mehmood, S., Rasul, S., Murtaza, I., & Khan, E. U. (2018). Dosimetric effect of limited aperture multileaf collimator on VMAT plan quality: A study of prostate and head-and-neck cancers. Reports of Practical Oncology and Radiotherapy, 23(3), 189-198. doi: 10.1016/j.rpor.2018.02.006. (Ghulam Murtaza (SHS), (JCR Listed))en_US
dc.identifier.urihttps://escholar.umt.edu.pk/handle/123456789/3693
dc.language.isoenen_US
dc.publisherReports of Practical Oncology and Radiotherapyen_US
dc.subjectVMAT, Treatment planning, Collimator rotation, Limited aperture MLCen_US
dc.titleDosimetric effect of limited aperture multileaf collimator on VMAT plan quality: A study of prostate and head-and-neck cancersen_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Dosimetric effect of limited aperture multileaf collimator on VMAT plan quality A study of prostate and head-and-neck cancers.pdf
Size:
5.95 MB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: